DPT-21_22 Programmable 2- wire dewpoint transmitter (-100..+20Cdp)

Product description

DPT-21 / DPT-22 – Programmable Gas Humidity Transmitters

DPT-21 and DPT-22 are modern, microprocessor-based transmitters designed for precise gas humidity measurement, expressed as dew point temperature, relative humidity, water vapor content (ppm), or degrees Fahrenheit. The devices convert the signal from a ceramic capacitive sensor into a standard 4–20 mA current signal.

The transmitters are available with two types of process connections:

  • DPT-21 with external thread 5/8″-18 UNF
  • DPT-22 with thread G1/2″

The transmitters come factory-calibrated and ready to use. However, key parameters – such as physical unit, measurement range, output signal type, and fault indication mode – can be adjusted by the user using the dedicated DPT-IF2013U USB interface.

Key Features:

  • Dew point temperature measurement range: –100 to +20 °Cdp
  • Configurable measurement span
  • Available units: °Cdp, °Fdp, ppm (v), %RH
  • Fast humidity response time (T95 ~1 min)
  • Robust stainless steel housing (IP65 protection)
  • Maximum working pressure: up to 40 MPa
  • Loop-powered: 8–36 V DC
  • Compatible with SCADA / PLC systems

Applications

With dew point measurement down to –100 °Cdp, the DPT-21 / DPT-22 transmitters are ideal for demanding environments where ultra-dry gas conditions are critical:

  • Semiconductor industry – humidity control during microprocessor production
  • Pharmaceutical industry – dry storage and transport of pharmaceuticals
  • Aerospace – protection of systems against freezing and corrosion
  • Energy and gas industry – monitoring gas quality and preventing condensation
  • Technical gas production – controlling purity and humidity during cylinder filling

DPT-21 and DPT-22 are an excellent choice for industrial applications where reliable and accurate gas humidity measurement is essential.

Technical data

 

Process connection external thread 5/8″-18 UNF
external thread G1/2″
Dewpoint measurement range -100÷20 °Cdp, -148÷68 °Fdp*, 0.5÷23000 ppm(v)
Measurement accuracy ±2 °Cdp
Thermal drift compensated across operating temperature range
Conversion range 10÷100 % of measurement range – programmable
Response time T95 1 min (dry to wet)
Output signal 4÷20 mA or 20÷4 mA – programmable
Linear region of output signal 3.8÷20.5 mA
Output signal delay after power on ca 5 s
Sensor failure indication 23 mA or 3,5 mA – programmable
Power supply Vs 8÷36 V DC / 24 mA (from current loop)
Maximal load resistance (Vs[V] – 8) / 0,024
Operating temperature range -40÷60 °C
Maximal operating pressure 40 MPa
Housing material / filte stainless steel / 5 µm stainless steel sintered guard
Ingress protection IP65
Dimensions with connector (diameter x length) / weight 31 x 129 mm / 150 g
Accessories (ordered separately interface IF-2013U-DPT
process connection adapter G1/2″-DPT
sampling block SB-DPT

Ask about the product

(+48) 22 720 31 80

Our experts will help you to customize the product